Main technical parameters and performance of CLJ-E dust particle counter:
1. Light source: semiconductor laser light source
2. Sampling volume: 2.83L/min (0.1cfm/min)
3. Detection range: 100 to 1 million levels
4. Allow the tested air to have a dust concentration of no more than 100000 particles/2.83L
5. Particle size channel: 0.3 0.5 1.0 3.0 5.0 10.0 (μ m) six levels
6. Display or print can convert particles contained within 2.83 liters/minute into particles contained within 1m3.
7. Sampling period: 1-10 (min)
8. Self purification time: ≤ 15 (min)
9. Calibration: Traceable to the National Institute of Standards and Technology (NIST) in the United States. Our company has passed the national metrology standard building assessment and can be traced back to the Institute of Metrology and Measurement Technology for self calibration or third-party national metrology institutions for calibration
10. Working environment: Temperature: 10-35 ℃ Relative temperature: 20-75% RH
11. Maximum power consumption: 25W
12. Range and accuracy of measuring temperature and humidity: (optional)
(1) Temperature: 0-50 ℃± 1 ℃
(2) Humidity: 0~* RH ± 5%
13. Sampling points set from 2 to 7 points
14. Set the sampling frequency to 2-9 times per point
UCL report: compliant with FS-209E and Chinese GMP standards
16. Working hours: 8 hours
17. Power supply: AC220V ± 10% 50 ± 2Hz
18. Weight: 4.8kg
19. Dimensions: 260 × 130 × 340